Control Input |
Number of commands : 3 Kinds : HOLD, HOLD RESET, BALANCE. Input system : No-voltage contact signal or open collector signal (should enable application of 12V DC and current flow of 5 mA.) |
Power Supply |
AC 100 to 240V±10%, 50/60 Hz, 20 VA or less |
Applicable Bridge Resistance |
87.5Ω to 10kΩ(Up to 4 transducers with 350Ωbridge resistance can be connected in parallel) |
Measuring Range |
±3.2 mV/V (including tare) (±6400μm/m) |
Adding Function |
Setting range ±19999 |
Sampling rate |
2000 times/sec. |
Bridge Excitation |
10V DC/2V DC, switchable by panel switch |
Analog (D-A) Output |
Setting items : Indication for 0 V output (±19999), Indication for 10 V output (±19999) Nonlinearity : Within ±0.1% FS |
Number of Measuring Channels |
1.0 |
Weight |
Approx. 1.1 kg |
Calibration function |
Possible by inputting a numeric value or by applying an actual load |
Applicable transducer |
Strain gage transducers |
Indicator |
±19999 (Decimal point can be put anywhere.) Character height 14 mm, red LED Mode : Normal/hold Update : 15.6 times/sec. in normal mode |
Analog (D-A) Output |
Current Output : 4 to 20 mA (load resistance 500Ωor less) Corresponds to voltage output of 0 to 10 V. |
Check Function |
Self-check |
Analog (D-A) Output |
Voltage Output : ±10 V (load resistance 2kΩ or more), Desired scaling possible |
High/low limit Comparative function |
Number of setting points : 2 (high limit, low limit), Setting range : ±19999, Max. Hysteresis width : Can be set in a range of 0 to 19999, Output : Open collector, Load capacity : 30 VDC, 20 mA (resistive load), Comparison mode : Normal/hold, Comparison rate : 500 times/sec. |
EMC Standard |
Conforms to IEC61326-1 (class A) |
Operating Temperature/Humidity Range |
-10 to 50°C, 20 to 85% RH (noncondensing) |
Safety Standard |
Conforms to IEC61010-1 (installation category II, pollution degree 2) |
Dimensions |
96(W)×96(H)×139(D)mm (excluding protrusions) |
Analog (D-A) Output |
Stability : Sensitivity stability : Within ±0.01%/°C Zero stability : ±150 μV RTI /°C |
Frequency Response Range |
DC to 100 Hz (within -3 dB at 100 Hz) DC to 10 Hz (within -3 dB at 10 Hz) |
Nonlinearity |
Within ±(0.03% FS + 1 digit) |
Panel Cut Dimensions |
92×92mm |
Original Value Measurement Function |
±3.200mV/V |
Digital Zero Adjustment Range |
Same as measuring range (digital adjustment only) |
Measurement Mode |
Time-based peak hold : Peak in a designated period is held. |
Input/Output Terminal Board |
M3 screw terminal (Applicable crimp-style terminal V1.25-3A, 1.25-3B or the equivalent) |
Analog (D-A) Output |
Resolution : 13 bits Conversion rate : 500 times/sec. Insulation withstand voltage between the DA-output and main circuit : 250 VAC for one minute (Circuit between voltage output and current output is not isolated.) |
Measurement Mode |
Point-based hold : Value measured at a selected point is held. Section-based peak hold : Peak in a designated section is held. Peak hold : Holds a maximum value during measurement. Normal : Measured value is normally indicated. |
Zero compensation function |
Automatic zero compensation : Uses digital zero within set range. Approximated zero compensation : Indication is made zero when the reading is in a preset range of 0 to 9. |
Smoothing Function |
Minimum scale : 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000 counts. Number of moving averaging times : 8, 16, 32, 64, 128, 256, 512, 1024, 2048. Analog filter : 10, 100 Hz |
Stability |
Zero Stability : within ±0.25 μVRTI / °C, Sensitivity Stability : within ±0.01%/°C |